Pulse rise time slow, Pulse width, Pulse rise time – HP 8360 User Manual
Page 296: Mod)

Pulse Rise Time
Function Group
Menu Map
Description
Programming Codes
See Also
MODULATION
4
This
lets you set the synthesizer to apply a slow rise pulse
filter to both internal and external pulse waveforms. This results in
pulses having approximately 2
rise/fall times. An asterisk next to
the key label indicates that this function is active.
SCPI:
PULM:SLEW
Analyzer:
NONE
Pulse Menu
Pulse Width
Function Group
Menu Map
Description
Programming Codes
See Also
(MOD)
4
This
lets you set a value for the internal pulse generator’s
pulse width. The range of acceptable values is from 1
to 65.5 ms.
The factory preset value is 1 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI:
PULSe:INTernal:WIDTh
Analyzer:
NONE
Pulse Menu
Operating and Programming Reference HP 8360