Pt trig menu, Pulse delay normal, Pt trig – HP 8360 User Manual
Page 288: Pulse normal
Pt Trig
Function Group
FREQUENCY
Menu Map
Description
This
accesses the list mode point trigger softkeys.
List
Pt Trig Auto
Automatically steps the synthesizer
to next point in the frequency list.
List Mode Pt Trig Bus
Steps the synthesizer to the next
point in the frequency list when an
HP-IB trigger is received.
List Mode Pt Trig Ext
Steps the synthesizer to the next
point in the frequency list when
an external hardware trigger is
received.
Programming Codes
SCPI:
NONE
Analyzer:
NONE
See Also
listed above, List Menu
Pulse
Normal
Function Group
Menu Map
Description
This
(Option 002 only) lets you set a value for the internal
pulse generator’s pulse delay. The output pulse is delayed from
the leading edge of the PULSE SYNC OUT signal. The range of
acceptable values is from 0 to a maximum of 25 ns less than the
period. The factory preset default is 0 ns. Use the numeric entry
keys, arrow keys, or rotary knob to change the value. When this
feature is active, its current value is displayed in the active entry
area.
Operating and Programming Reference HP 8360
User’s Handbook