Entry on/off, Ext det cal – HP 8360 User Manual
Page 202
Ext Det Cal
Function Group
Menu Map
Description
Programming Codes
See Also
ENTRY
NONE
This
lets you turn off (blank) the active entry area and
disable the ARROW keys, rotary knob, and entry keys. When
any function key (hard or soft) is pressed, the active entry area is
reactivated. The yellow LED, ENTRY ON, next to
indicates whether the entry area is active (LED on=active).
SCPI:
No specific code activates
Arrow Keys
“Entry Area” in Chapter 1.
Ext Det Cal
Function Group
USER CAL
Menu Map
Description
This
enables the synthesizer to act as a controller to an HP
power meter. This
causes an immediate execute on
the interface bus and generates an HP-IB error if no power meter is
present on the interface bus or if the synthesizer is unable to address
the power meter. Use external detector calibration to characterize
and compensate for an external negative diode detector used in an
external leveling configuration.
Programming Codes
CALibration:PMETer:DETector:INITiate?
CALibration:PMETer:DETector:NEXT?
suffix]
See Also
“Optimizing Synthesizer Performance” in Chapter 1.
HP 8360
User’s Handbook
Operating and Programming Reference