beautypg.com

Programming codes, See also – HP 8360 User Manual

Page 363

background image

Step Control Master

SCRLAR

SLRVE

S Y N T H E S I Z E R S Y N T H E S I Z E R

Figure S-l. Connections Required for a Two-Tone Scalar Network Analyzer Measurement System

1. Designate one synthesizer as the master, the other as the slave.

2. Make the connections.

3. To avoid synchronization problems, always set up the slave

(frequency and power) before setting up the master.

4. Set up the master (frequency, power, and sweep time).

5. Set the sweep time on the slave.

6. Configure the synthesizers for step sweep, or ramp sweep.

7. Select the appropriate triggering scheme.

8. Activate the slave mode on the slave synthesizer.

9. Activate the master mode on the master synthesizer.

By connecting the master’s 10 MHz reference standard to the slave’s

10 MHz reference input, the master synthesizer’s

supplies

the frequency reference for both synthesizers.

In step sweep measurements, if the master synthesizer is not
connected to an external controller, it must automatically trigger
the steps. Set

Step

Pt Trig Auto on the master.

When

a the scalar network analyzer is the step sweep controller, set

Step

Trig Bus on the master synthesizer so that the

analyzer can trigger the steps.

Programming Codes

SWEep:CONTrol:STATe
SWEep:CONTrol:TYPE

Analyzer: NONE

See Also

Step Control Slave, Step Swp

Menu

HP 8360

User’s Handbook

Operating and Programming Reference