Sweep span cal always, Swp span cal once, Cal always – HP 8360 User Manual
Page 373
![background image](/manuals/89476/373/background.png)
Swp
Span
Cal
Always
Function Group
USER CAL
Menu Map
Description
This
causes a sweep span calibration each time the frequency
span is changed. An asterisk next to the key label indicates this
feature is active.
Programming Codes
SCPI: CALibration:SPAN:AUTO
Analyzer: NONE
See Also
Freq Cal
Menu
“Using Frequency Calibration,” in Chapter 1.
Swp Span Cal
Once
Function Group
USER CAL
Menu Map
Description
This
activates sweep span calibration immediately and
performs it only once. An asterisk next to the key label indicates this
feature is active.
Programming Codes
SCPI: CALibration:SPAN[:EXECute]
Analyzer: NONE
See Also
Freq Cal Menu
“Using Frequency Calibration,” in Chapter 1.
HP 8360
User’s Handbook
Operating and Programming Reference