beautypg.com

Altrnate regs, Am bw cal always – HP 8360 User Manual

Page 157

background image

ALC

to remain there for all sweep and
modulation conditions.

See Also

“Optimizing Synthesizer Performance” in Chapter 1.

Function Group

SYSTEM

Menu Map

Description

This

causes the synthesizer to alternate on successive sweeps

between the present instrument state and a second instrument state
stored in an internal register (1 to 8). Select

Altrnate Regs once to

turn it on, a second time to turn it off. An asterisk next to the key
label indicates that this feature is active.

Programming Codes

SYSTem:ALTernate:STATe
SYSTem:ALTernate

Analyzer:

where

1 through 8 function on,

function off

See Also

[RECALL),

“Saving and Recalling an Instrument State” in Chapter 1.

AM BW Cal Always

Function Group

Menu Map

Description

Programming Codes

See Also

USER CAL

9

This

causes an AM bandwidth calibration to be performed

every time a frequency or power parameter is changed.

SCPI:

CALibration:AM:AUTO ON

Analyzer:

NONE

Modulation

Operating and Programming Reference

HP 8360

User’s Handbook