Factory

Preset Mode

Factory

Function Group

Menu Map

Description

Programming Codes

See Also

SYSTEM

8

This

sets the standard starting configuration of the

synthesizer when the

key is pressed, as set by the

manufacturer. An asterisk next to the key label indicates that this

feature is active. The following is a description of the configuration.

n

Start sweep at the minimum specified frequency.

Stop sweep at the maximum specified frequency.

n

Power level set at 0

n

Sweep time set to auto.

n

CONT sweep.

Sweep mode ramp.

n

ALC leveling point internal.

n

ALC leveling mode normal.

n

Markers set to activate at the center frequency of the sweep.

n

All function values stored in memory registers 1 through 9 remain

in their previous states.

n

The checksum of the calibration data is calculated, and if an error

is detected, the calibration data in protected memory is used. If
the checksum of the protected data is not correct, then default

values are used an error message (EEROM FAILED, LOST CAL)
is displayed.

SCPI:

SYSTem:PRESet:TYPE

Analyzer:

IP, which is the same as

P R E S E T

Preset Mode User

“Changing the Preset Parameters,” in Chapter 1.