3 meas upol - measuring parameters, Measuring parameters – Metrohm tiamo 2.4 (process analysis) User Manual
Page 999

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
983
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 867
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos, 855 and 867.
5.6.4.6.3
MEAS Upol - Measuring parameters
Tab: Method
▶ MEAS Upol ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
The two measuring modes Measurement with drift control or Mea-
surement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control