6 sample changer - heater/gas, Heater/gas – Metrohm tiamo 2.4 (process analysis) User Manual
Page 1703

■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1687
Selection
On/Off | Duration
Default value
On/Off
Duration
Entry of the duration during which the pump/valve should remain
switched on. This field is only displayed if Operation = Duration.
Input range
1 - 999,999 s
Default value
60 s
[Start]
Starts pump 2. The time elapsed since the start is shown in the status dis-
play.
[Stop]
Stops pump 2.
8.4.4.6
Sample changer - Heater/Gas
Program part: Manual
The heater and gas flow on the 774 and 874 Oven Sample Processor can
be switched on and off here.
Heater
Heating the oven up to the desired temperature during the defined heat-
ing time.
Temperature
Temperature to which the oven has to be heated. With Init it is heated up
to the initial temperature set on the instrument.
Input range
50 - 250 °C
Default value
50 °C
Selection
Init
Heating period
Duration of the heating period until the desired temperature is reached.
Input range
1 - 999 min
Selection
off
Default value
off
[Start]
Starts the heater. The current temperature is shown in the status display.
[Stop]
Stops the heater.