7 Devices
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1637
Unit
Unit of the measured value to be used.
Only editable for controlled variable = calculated value, otherwise
grayed out and not available for editing.
Entry
50 characters
Default value
% or controlled variable unit, respectively
Default value
Selection or input of the default value which is to be used for the input if
the result is invalid.
Input range
0% mapping of the input - 100% mapping of
the input
Selection
last value | Setpoint value
Default value
last value
Setpoint value
Setpoint value to which the control is to be set.
Input range
0% mapping of the input - 100% mapping of
the input 'Mapping unit of the input or of the
calculated value'
Default value
0 'Mapping unit of the input or of the calcu-
lated value'
Hysteresis
Hysteresis for setpoint value.
Input range
0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the analog input or of the calcu-
lated value'
Default value
0 'Mapping unit of the analog input or of the
calculated value'
Control value
Selection of the control value from all of the digital outputs of the Con-
troller output type.
Selection
'Port designation'
'Port designation'
Only digital output ports of the Controller output type can be
selected.