2 General program functions

tiamo 2.4 (for Process Analysis)

■■■■■■■■

49

Identifica-
tion

Description

Commands

.OVS

Value of the output at the start of
the command.

Direct controller/2-point control-
ler command status: 1 = active,
0 = inactive

PID/PWM: -100 to 100% (value of
the analog output or the pulse
frequency)

CONTROL

.PK{x}.ABS

Absorbance of the peak x (1 - 9)
in mAU

MEAS Spec

.PK{x}.TRN

Transmission of the peak x (1 - 9)
in %

MEAS Spec

.PK{x}.CNT

Intensity of the peak x (1 - 9) in
counts

MEAS Spec

.PK{x}.CNR

Intensity of the peak x (1 - 9) in
the reference spectrum in counts

MEAS Spec

.PK{x}.CND

Intensity of the peak x (1 - 9) in
the dark spectrum in counts

MEAS Spec

.PK{x}.WVL

Wavelength of the peak x (1 - 9)
in nm

MEAS Spec

.PK{x}.SAT

Detector with peak x (1 - 9) satu-
rated: 1 = saturated;0 = not
saturated

MEAS Spec

.RAN

Current absolute rotation angle of
the rack in ° in relation to the axis
of the selected tower (entry when
exiting the command)

MOVE

.RE{x}.DRC

Correlation coefficient for mean
dosing rate in window x (1 - 9)

STAT

.RE{x}.DRS

Standard deviation for mean dos-
ing rate in window x (1 - 9) in
mL/min

STAT

.RE{x}.DRT

Mean dosing rate in window x (1
- 9) in mL/min

STAT

.RE{x}.RWL

Lower limit of the evaluation win-
dow x (1 - 9) in s

STAT