Metrohm tiamo 2.4 (process analysis) User Manual
Page 66

2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Identifica-
tion
Description
Commands
.RE{x}.RWH
Upper limit of the evaluation win-
dow x (1 - 9) in s
STAT
.RES
Calculated result of the standard
addition in the selected unit
STDADD
.RPO
Current rack position (entry when
exiting the command); 0 means
'not defined'
MOVE
.RTE
Reference temperature in °C
MET Cond, MEAS
Cond, CAL Cond
.SAN
Current absolute swing angle of
the robotic arm in ° (entry upon
exiting the command)
MOVE, SWING
.SLO
Electrode slope of the sensor used
for the command (in % for pH
sensor or mV for ISE sensor) or
electrode slope calculated from
the calibration (for STDADD and
CAL LOOP)
DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS
T, MEAS Conc,
STDADD, CAL
LOOP, DOS, ELT
LOOP
.SME
Start measured value (measured
value after processing the start
conditions) in the unit of the
measured value
DET, MET, SET,
KFT, KFC, BRC,
STAT
.SPO
Current external position (entry
when ending the command); 0
means invalid position
SWING
.STC {x}
Concentration of the standard x
(1 - 50) used for the command
CAL LOOP Opt
.STE
Start temperature (temperature
after processing the start condi-
tions) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT