5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
763
Command variables
The following command variables are generated by the SET Upol com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CBY
Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated
.COK
Command status; 1 = Condition requirement ful-
filled, 0 = Condition requirement not fulfilled
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in
s
.DRI
Current and/or last drift for drift correction in µL/min
.DSC
Time for processing all start conditions in s
.DTI
Time in s for drift correction (time from the start of
the titration to the end of the command)
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.EP{x}.MEA
Measured value for the endpoint × (1 - 9) in the unit
of the measured value
.EP{x}.MEP
Number of endpoints in the window × (1 - 9); 1 = 1
endpoint, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift
.EP{x}.TEM
Temperature for the endpoint × (1 - 9) in °C
.EP{x}.TIM
Time in s until the endpoint × (1 - 9) is reached
.EP{x}.VOL
Volume for the endpoint × (1 - 9) in mL
.ETE
End temperature (temperature after the command
has been processed) in °C