5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1109
Data transmission from standard
Input of the data of the conductivity standard, which should be as close as
possible to the Current temperature. This data is as a rule contained in
the Data Sheet/Certificate for the conductivity standard.
Reference temperature
Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.
Input range
-20.0 - 150.0°C
Default value
20.0
Conductivity
Input of the conductivity of the standard at the reference temperature.
Input range
0.00000 - 2000.00 mS/cm
Default value
1.0 mS/cm
Temperature coefficient
Temperature coefficient of the standard with the reference temperature.
Input range
0.00 - 9.99 %/°C
Default value
1.00 %/°C
5.6.5.5
CAL Spec
5.6.5.5.1
CAL Spec - Overview
Dialog window: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command
name'
Command for wavelength calibration.
Devices
This command can be executed with the following instrument:
Spectrometer: Avantes