5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
888
■■■■■■■■
tiamo 2.4 (for Process Analysis)
[New]
Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint.
[Properties]
Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.4.3.10, page 717).
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.1 - 20.0 pH/mL
Default value
1.0 pH/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.1 - 20.0 pH/mL
Default value
1.0 pH/mL