5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
547
Initial measured value
NOTE
Only displayed for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
Drift for the measured value acceptance during the titration.
Titrando, 855
Input range
0.01 - 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Titrino
Input range
0.05 - 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.