5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
531
5.6.3.2.4.7
DET Ipol - Additional evaluations
Tab: Method
▶ DET Ipol ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint (see Chapter 5.6.3.2.4.11, page 538).
[Properties]
Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.2.4.11, page
538).
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.