5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
849
Temperature measurement
Type of temperature measurement.
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Cell
Generator type
Selection of the generator type.
Selection
without diaphragm | with diaphragm
Default value
with diaphragm
Generator current
Polarization current at the generator electrode.
Selection
100 mA | 200 mA | 400 mA | auto
Default value
auto
auto
Means that the current is automatically adapted to the conductivity of
the reagent and that near the endpoint the current will be controlled
at smaller values.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.