5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
840
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Selection
automatic | manual | off
Default value
automatic
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Input range
0.0 - 99.9 µg/min
Default value
0.0 µg/min
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Input range
1 - 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Input range
0 - 999 s
Default value
0 s
Automatic start after sample addition time
on | off (Default value: on)
If this option is activated, then the titration will be started automatically
after the sample addition time.
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Input range
0 - 99 s
Default value
0 s