3 properties - titer, Titer – Metrohm tiamo 2.4 (process analysis) User Manual
Page 1420
6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1404
■■■■■■■■
tiamo 2.4 (for Process Analysis)
[E-mail...]
[E-mail...] opens the Send e-mail window (see Chapter 2.6.1, page 93).
Acoustic signal
on | off (Default value: off)
If this option is enabled, an acoustic signal will be emitted in addition to
the message defined above.
Action
The actions can only be edited if monitoring is switched on.
If during monitoring it is found that the working life has expired then one
of the following actions will be triggered automatically during the start
test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the working life has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, the message that the
working life has expired is automatically saved with the determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
6.6.3.3
Properties - Titer
Tab: Configuration
▶ Titrants/Solutions ▶ [Edit] ▶ Properties... ▶ Titer
Titer determination
Titer (value)
Titer value of the solution. The value for the titer of a solution used by the
titration commands DET, MET, SET and KFT is available for calculations
as a variable 'Command name.TITER'.
Input range
0.00000 - 9.99999999E+9 (max. 10 digits)
Default value
1.000
Titer (unit)
Titer unit of the solution.
Entry
10 characters