Page 1467

■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1451
■
Limit values
Definition of the limit value monitoring for outliers and the coefficient
of determination.
■
History
Display of the last 10 calibration values.
6.8.3.2
Properties - Colorimetric sensors
Tab: Configuration
▶ Colorimetric sensors ▶ Edit ▶ Properties... ▶ Colorimetric
sensor
Name
Name of the sensor.
Entry
24 characters
Comment
Possibility to enter remarks about the sensor.
Entry
24 characters
Default value
'empty'
6.8.3.3
Properties - Calibration data
Tab: Configuration
▶ Colorimetric sensors ▶ Edit ▶ Properties... ▶ Calibration
data
Wavelength
Wavelength at which the calibration of the instrument has been carried
out.
Input range
100.0 - 2000.0 nm
Default value
400.0 nm
c0
Calibration coefficient of zero order.
Input range
-1E+99 - 1E+99 (max. 10 significant places)
Default value
0
c1
Calibration coefficient first-order.
Input range
-1E+99 - 1E+99 (max. 10 significant places)
Default value
0
c2
Calibration coefficient second-order.
Input range
-1E+99 - 1E+99 (max. 10 significant places)
Default value
0