Metrohm tiamo 2.4 (process analysis) User Manual
Page 61
![background image](/manuals/586371/61/background.png)
■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
tiamo 2.4 (for Process Analysis)
■■■■■■■■
45
Identifica-
tion
Description
Commands
.IVS
Value of the input at the start of
the command.
CONTROL
.LCO
Loop counter = current number
of completed runs, both of
Repeat loops and of While loops
LOOP, CAL
LOOP, ELT LOOP
.LP.CAx
Calculated value x (1 - 3) for the
last measuring point on the mea-
suring point list
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without Ref,
Spec, Opt), DOS
.LP.CHA
Charge for the last measuring
point of the measuring point list
in mA s
KFC, BRC
.LP.CNT
Intensity of the last measuring
point of the measuring point list
in counts
MEAS Opt, MEAS
Opt Conc, CAL
MEAS Opt
.LP.DME
Measured value differential for
the last measuring point on the
measuring point list
MET
.LP.DVT
dV/dt for the last measuring point
on the measuring point list (SET,
KFT, STAT, DOS) or drift for the
last measuring point on the mea-
suring point list in µg/min (KFC,
BRC)
SET, KFT, KFC,
BRC, STAT, DOS
.LP.ERC
ERC for the last measuring point
in the measuring point list
DET
.LP.EXx
External value x (1 - 3) for the last
measuring point on the measur-
ing point list
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without Ref,
Spec, Opt), DOS
.LP.GFL
Gas flow value for the last mea-
suring point on the measuring
point list in mL/min
MEAS T/Flow
.LP.IGE
Current pulsed current for the last
measuring point in the measuring
point list in mA
KFC, BRC