5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
800
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Stop time
Stop when the entered time has elapsed after the start of the titration
(including start conditions).
Titrando
Input range
0 - 999,999 s
Selection
off
Default value
off
Filling rate
Rate at which the dosing cylinder is to be filled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing unit used.
Titrando
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.5.2.7
KFT Ipol - Conditioning
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Entry
25 characters
Conditioning
Conditioning
on | off (Default value: off)
Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift
is lower than the Start drift entered here. In such cases, the Status dis-
played in the Run subwindow on the Single determination or Deter-
mination series tab will then switch to COND READY.