6 Configuration
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1405
Selection
mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | ppm | % | mEq/L
Default value
mol/L
Date titer det.
Shows the date and time of last titer determination, which are automati-
cally entered with each automatic titer determination or manual entry.
Titer method
Name of the method with which the last titer determination was carried
out. If the titer has been entered manually then manual will be shown
here.
User
Short name of the user logged in during titer determination or when the
titer was entered manually. If the login function is not used, then the user
logged in under Windows will be entered automatically.
Statistics
Statistics data will only be shown if the titer has been assigned by the titer
method as a mean value of a multiple determination. If the titer has been
determined with a single determination or entered manually then no sta-
tistics values will be shown.
Number of titer determinations (n)
Number of determinations used to determine the titer.
Abs. standard dev. (s abs)
Absolute standard deviation for the titer determination in the unit of the
titer.
Rel. standard dev. (s rel)
Relative standard deviation for the titer determination in %.
Monitoring of titer validity
on | off (Default value: off)
If this option is activated, then the validity of the titer will be monitored.
Validity of titer
Validity of the titer in days. If a value is entered here, then the date for the
Next titer determination will be automatically adapted.
Input range
0 - 999 days
Default value
999 days