6 Configuration
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1431
Calibration date
Date and time of last calibration that is entered automatically after each
automatic calibration or manual entry (is not being displayed for the Ther-
moprobe).
Calibration method
Name of the method with which the last calibration was carried out (is
not being displayed for the Thermoprobe). If the calibration data has been
entered manually, then manual will be shown here.
User
Short name of the user logged in during the calibration or when the cali-
bration data was entered manually (is not being displayed for the Ther-
moprobe). If the login function is not used, then the user logged in under
Windows will be entered automatically.
Measuring input
Display of the type and the serial number of the measuring input at which
the sensor was connected during the calibration (only for Titrando, is not
being displayed for the Thermoprobe).
[Initial data]
Opens the Initial data Sensor dialog window (see Chapter 6.7.3.7,
page 1440), in which the initial calibration data stored in the sensor chip
is displayed (only for intelligent sensors).
Calibration data monitoring
on | off (Default value: off)
If this option is activated, then the validity of the calibration will be moni-
tored (is not being displayed for the Thermoprobe).
Calibration interval
Validity of the calibration in days. If a value is entered here, then the date
for Next calibration will be automatically adapted.
NOTE
The days are monitored, i.e., the calibration becomes invalid when the
date changes and not at the time when the calibration took place.
Only editable if monitoring is switched on.
Input range
0 - 999 days
Default value
999 days