beautypg.com

5 special features, 1 crystal switching – INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 75

background image

3 - 25

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

3.5 Special Features

The Cygnus 2 has several special features to enhance the performance of the
instrument.

3.5.1 Crystal Switching

The Cygnus 2 offers a choice of Single, XtalTwo (CrystalTwo®), XtalSix
(CrystalSix®), Xtal12 (Crystal12®), or Generic sensors. The CrystalTwo,
CrystalSix, Crystal12, and Generic sensors provide one or more backup crystals in
case a crystal fails during deposition. Sensor types are specified on the Sensors
screen.

The XtalTwo option requires the 779-220-G1 or 779-220-G2 XTAL2 Switch. The
XTAL2 Switch is connected to any Sensor input through an XIU package.

All multi-position and shuttered sensors require the Pneumatic Actuator Control
Valve, part number 750-420-G1, and a feedthrough with an air tube.

A crystal switch will automatically occur when:

The Cygnus 2 is configured for a XtalTwo (CrystalTwo) sensor type, a Material
is STARTed or running and there is another good crystal available when the
active crystal fails.

The Cygnus 2 is configured for a XtalSix, Xtal12 or Generic, a Material is
STARTed or running, and there is at least one good crystal left in the carousel
when the active crystal fails.

Using a XtalSix or Xtal12 and pressing START if the current position is not
within the First/Last Xtal position range listed on the Sensor page of the
Material screen.

S & Q levels are exceeded while in deposit.

A crystal switch will NOT automatically occur:

20. IDLE

Source resting at Idle Power; will accept a
START command.

Inactive

Inactive

19

21. STOP

The respective Material’s Source output is
set to zero power. The Material’s display is
frozen at the last thickness value.

Inactive

Inactive

20

NOTE: In STOP the Cygnus 2 will accept a START provided a valid crystal is available for the Material being
started.

Table 3-2 State descriptions (continued)

STATE

CONDITION

RELAY CONTACT

STATUS

REMOTE

COMM

ENCODES

Source
Shutter

Sensor
Shutter