beautypg.com

INFICON Cygnus 2 Thin Film Deposition Controller User Manual

Page 191

background image

9 - 43

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

11

(0x0b)

Zero Counter

Counter Range is 0-20

0 = All

12

(0x0c)

Zero Timer

Timer Range is 0-20

0 = All

13

(0x0d)

Start Timer

Timer Range is 1-20

14

(0x0e)

Cancel Timer

Timer Range is 0-20

0 = All

15

(0x0f)

Clock Hold On

Stops timer in Pre-Deposit or Post-Deposit. Keeps in

current state at current power setting.

Material Range 0-6 (0 = all Materials)

16

(0x10)

Clock Hold Off

Restarts timer in Pre-Deposit or Post-Deposit

Material Range 0-6 (0 = all Materials)

17

(0x11)

User message on

Message Range is 1-10

18

(0x12)

User message off

Message Range is 1-10

19

(0x13)

Soak Hold 1 on

Material Range is 0-6

(0 = all Materials)

20

(0x14)

Soak Hold 1 off

Material Range is 0-6

(0 = all Materials)

21

(0x15)

Soak Hold 2 on

Material Range is 0-6

(0 = all Materials)

22

(0x16)

Soak Hold 2 off

Material Range is 0-6

(0 = all Materials)

23

(0x17)

Switch

Sensor

to

Sensor Range = 1 - 6

Crystal Range = 0 to maximum allowed crystals for the

requested Sensor. If Crystal # = 0, will switch to the "next"

crystal. If the Sensor is not used in a Material currently

running, the next crystal will be the next one in the

sequence. If the Sensor is used in a Material that is

running, the next crystal will be the next one from the XL

Position parameters sequence. If a Sensor is used in a

Material that is running, the value of "Crystal #" must be

zero, otherwise, an error will occur.

Table 9-15 Remote general (continued)

RG

Command

ID

Meaning

Action Value

Notes