13.5 Input Error Messages
Keyboard Input Error Messages come up on the entry of invalid parameters. These
messages will appear on the transient error message line in the message box. The
font color of the parameter value being updated will change to a different color to
denote an input error has been made. They will clear when the clear key is pressed.
The original stored value will then be re-displayed. See
.
Table 13-5 Input error messages
Input Error Message
Set
Cleared
Screen
Value Too High
The value entered is larger than the maximum
value allowed.
Each input error
message is
taken down by
pressing the
clear key.
Any Parameter
edit
Value Too Low
The value entered is lower than the minimum
value allowed.
Any Parameter
edit
Duplicate Source
Can use each Source number only once for
thickness sum.
General
Invalid Mask
An attempt was made to enter an incorrect
Ethernet net mask.
General
Incorrect Time
An attempt was made to enter more then 59
seconds or 99 minutes into a time parameter.
Material
Ramp 1 > Ramp 2
Invalid Material parameter ‘Ramp’, values were
entered. Start Ramp 1 thickness value must be
less then Start Ramp 2 value when the Start
Ramp 2 value is non-zero.
Material
Ramp 2 < Ramp 1
Invalid Material parameter ‘Ramp’, values were
entered. Start Ramp 2 thickness value must be
greater then Start Ramp 1 value.
Material
DAC Used Material ##
Or
DAC Used Sensor ##
The DAC number has been previously defined
as a recorder output in the Material or the Sensor
set up.
Source
parameter
DAC Used Source XX
The DAC number has been previously defined
as a Source DAC output in one of the Source set
ups
Material, Sensor
and Source
parameter
Input in use
An attempt was made to enter an input that is
already reserved for use in Source setup or one
of the logic statements.
Source
parameter
Output in use
An attempt was made to enter an output that is
already reserved for use. The output is already
reserved for use in Sensor or Source set up or in
one of the logic statements.
Sensor, Source
parameter and
Logic Edit
Invalid Lock Code
An incorrect lock code was used to attempt to
unlock the Cygnus 2.