9.4.31 Remote General Action
Command = RG
Command ID =
Action Value =
See
for commands that require an action value.
Response = None (Just header and trailer)
Table 9-15 Remote general
RG
Command
ID
Meaning
Action Value
Notes
1 (0x01)
Stop All
Duplicates the equivalent front panel action
2 (0x02)
Reset All
Duplicates the equivalent front panel action
3 (0x03)
Remote Lock On
No parameters can be updated from the front panel
"R Lock" displayed
4 (0x04)
Remote Lock Off
Cancel RG 3 command
5
(0x05)
Set Logic
Statement
Number> to True Range is 1-100. Only does something if logic statement event "Computer Control" set. Otherwise, ignored. 6 (0x06) Clear Logic Statement Number> to False. Range is 1-100. Only does something if logic statement event "Computer Control" set. Otherwise, ignored. 7 (0x07) Crystal Fail Inhibit On Range is 0-6 (0 = All) "Crystal Fail" output always set to FALSE, even during crystal fail 8 (0x08) Crystal Fail Inhibit Off Range is 0-6 (0 = All) "Crystal Fail" output has normal meaning 9 (0x09) Set date (Month, Day, Year) or (Day, Month, Year) Depending on General parameter 17 (Date Format). Range as follows:Month = 1 to 12 Day = 1 to 31 Year = 2000 to 2099 10 (0x0a) Set time (Hour, Minute, Second) Range as follows Hour = 0 to 23 Minute = 0 to 59 Second = 0 to 59