14 query material name, 15 update material name, 16 query input name – INFICON Cygnus 2 Thin Film Deposition Controller User Manual
Page 173: 17 update input name, 18 query output name

9 - 25
IP
N 07
4-
54
5-
P1
E
Cygnus 2 Operating Manual
9.4.14 Query Material Name
Command = QN
Material Number =
Response = < Material Name>
Material Name =
will be a blank if the default Material # name has not been changed.
9.4.15 Update Material Name
Command = UN < Material Number> < Material Name>
Material Number =
Material Name =
Response = None (Just header and trailer)
9.4.16 Query Input Name
Command = QI
Input Number =
Response =
Input Name =
9.4.17 Update Input Name
Command = UI
Input Number =
Input Name =
Response = None (Just header and trailer)
Input Names may only be updated if the Cygnus 2 is in Ready or Stop.
NOTE: UI changes are not allowed while a Material is running. The UI command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name.
9.4.18 Query Output Name
Command = QO
Output Number =
Response =
Output Name =