6 - 4

IP

N 07

4-

54

5-

P1

E

Cygnus 2 Operating Manual

If the Ethernet option is installed, the following parameters will be shown

IP ADDRESS . . . . . . . . . . . . . . . . . . nnn.nnn.nnn.nnn

Use the up and down cursor keys to scroll through the address and the
numerics keys to enter the desired address. The default address is
10.211.72.203. This setting is not cleared on pressing CLEAR on power-up.

NET MASK . . . . . . . . . . . . . . . . . . . nnn.nnn.nnn.nnn

The net mask is some number of leading bits set to 1, followed by all 0. The
default net mask value is 255.255.0.0: This setting is not cleared on pressing
CLEAR on power-up.

255

.255

. 0

.0

11111111 1111111 00000000 00000000

Test RS-232 Function Key

Once the cursor is in the parameter area, the Test RS-232 function can be
performed by pressing the F1 function key. Plug the loop-back connector
760-406-P1 supplied in the Cygnus 2 ship kit into the RS-232 connector on the
rear panel and press F1. Message RS-232 Test Failed or RS-232 Test Passed
will appear. If the loop-back connector is not available, jumper pins 2 and 3 to
perform the test. Contact INFICON Service if the RS-232 Test Failed message
appears.