8 remote communications, 9 accessories, 10 power – INFICON Cygnus 2 Thin Film Deposition Controller User Manual
Page 33: 11 operating environment, 12 storage temperature, 13 warm up period

1 - 15
IP
N 07
4-
54
5-
P1
E
Cygnus 2 Operating Manual
1.4.8 Remote Communications
RS232C Serial Port . . . . . . . . . . . . . Standard; INFICON binary protocol
Baud Rates. . . . . . . . . . . . . . . . . . . . 115,200, 57,600, 38,400, 19,200, 9,600
Ethernet . . . . . . . . . . . . . . . . . . . . . . Optional, Programmable IP Address and
Net Mask
USB port. . . . . . . . . . . . . . . . . . . . . . Print screen bitmap (.bmp) files, Datalogging
file, Configuration file store/retrieve
1.4.9 Accessories
Connector Kit . . . . . . . . . . . . . . . . . . Connectors for inputs and relays
Operating Manual. . . . . . . . . . . . . . . 074-5000-G1 CD-ROM
1.4.10 Power
Power . . . . . . . . . . . . . . . . . . . . . . . . 100-230 V (ac) +/-15% V (ac); 50/60 Hz
+/-3Hz Maximum apparent power 150 VA
Fuse . . . . . . . . . . . . . . . . . . . . . . . . . use 250 V 4 Amp, 5 x 20, Slow Blow
1.4.11 Operating Environment
Usage . . . . . . . . . . . . . . . . . . . . . . . . Indoor only
Temperature . . . . . . . . . . . . . . . . . . . 0 to 50°C (32-122°F)
Humidity . . . . . . . . . . . . . . . . . . . . . . Up to 85% RH, non-condensing
Altitude . . . . . . . . . . . . . . . . . . . . . . . Up to 2000 meters
Installation (Overvoltage) . . . . . . . . . Category II per IEC 60664
Pollution Degree. . . . . . . . . . . . . . . . 2 per EN 61010
1.4.12 Storage Temperature
Storage Temperature . . . . . . . . . . . . -10 to 60°C (14 to 140°F)
1.4.13 Warm Up Period
Warm Up Period . . . . . . . . . . . . . . . . None required;
For maximum stability allow 5 minutes.