beautypg.com

GW Instek GSP-830 User Manual User Manual

Page 93

background image

LIMIT

LINE

93

No.

MHz

dBm

1

2

3

4

5

100

-40

102

98

-30

-40

No.

MHz

dBm

1

2

3

4

5

-40

102

98

-40

3. To undo deletion, press F5

(Undelete).

Undelete

F 5

3c. Delete the whole

limit line data

1. Press F4 (Delete All).

Delete All..

F 4

2. Press F1 (No) or F2 (Yes) to

confirm deletion. All 10 limit
line points will be deleted at
once.

No

F 1

Yes

F 2

3. Press F6 (Return) to go back to

the previous menu.

Return

F 6

4. To undo deletion, press F5

(Undelete).

Undelete

F 5

4. Switch

upper/lower
limit line

If necessary, press F1 (Limit) to
start editing the other limit line.
Repeat the above steps.

Limit

Low High

F 1

Example

High Limit Line

Point No. Frequency (MHz)

Amplitude (dBm)

1

98MHz

−40dBm

2

100MHz

−30dBm

3

102MHz

−40dBm

Resulted

line