beautypg.com

Set input impedance, Select input impedance (50ω/75ω), Set impedance offset (75ω only) – GW Instek GSP-830 User Manual User Manual

Page 57: Press f6 (more), Make sure 75ω is selected in f1 (input z), Press f2 (input z cal)

background image

AMPLITUDE

57

Set Input Impedance

Select input impedance (50Ω/75Ω)

Background

In most cases, the default 50Ω is appropriate. Use 75Ω
when specifically required, such as in cable TV signals.

1. Press the Amplitude key.

Amplitude

2. Press F6 (More).

More

F 6

Panel operation

3. Press F1 (Input Z 50Ω/75Ω) to

select the impedance.

4. When 75Ω is selected, the

amplitude icon appears at the
bottom of the display.

Set impedance offset (75Ω only)

Background

Impedance transformation to 75 Ω is also available
through external devices such as impedance converter
module (optional accessory ADP-101). In these cases an
external loss will be induced. Impedance offset can
compensate this effect.

1. Press the Amplitude key.

Amplitude

2. Press F6 (More).

More

F 6

Panel operation

3. Make sure 75Ω is selected in F1

(Input Z).

4. Press F2 (Input Z Cal).

Input Z Cal

5.9dB

F 2

5. Enter the offset using the

Numerical keys and dB or Enter
key.

BK SP

dB

Enter

GHz /

Sec

mSec

MHz /

kHz /

Sec

μ

Range

−10dB to +10dB, 0.1dB resolution