Page 128: Amp uncal, Med unlock, Ext unlock, Lo1 unlock, Lo3 unlock, Ref unlock

background image

GSP-830 User Manual

128

3. Press F2 (Self Test).

Self Test...

F 2

4. The GPIB result appears at F1.

If the result is Fail (underlined),
contact service personnel.

GPIB

Pass Fail

F 1

GPIB constraints

Keep these rules when using the GPIB interface.

Altogether less than 15 devices & 20m cable length,
2m between each device on the bus

Unique address assigned to each device

At least 2/3 of the GPIB devices turned on

No loop or parallel structure allowed

View System Information

View system error

Panel operation

View the bottom of the screen, the error message area.
If there is a system error, the message appears in red.

Center : 1.5GHz

(EXT Unlock)

EXT

Unlock

Amp

Uncal

Inappropriate RBW or VBW is selected.

Frequency is less than 15MHz and
amplitude is less than −30dBm.

EXT

Unlock

External reference input is not working
properly.

LO1

Unlock

Local oscillator 1 is not working properly.

LO3

Unlock

Local oscillator 3 is not working properly.

Med

Unlock

±

1ppm stability signal is not working

properly. Appears only when the optional
±

1ppm Stability Module is installed.

Ref

Unlock

Internal reference signal is not working
properly.