beautypg.com

Set vertical scale, Set reference amplitude, Press the amplitude key – GW Instek GSP-830 User Manual User Manual

Page 49: Press f1 (ref.level), See page 50 for se lecting amplitude units

background image

AMPLITUDE

49

Set Vertical Scale

Vertical display scale is defined by the reference
amplitude, amplitude range, measurement unit, and
external gain/loss.

Set reference amplitude

The reference level defines the amplitude at the top of the displayed range
(see below screen snapshot).

Panel operation

1. Press the Amplitude key.

Amplitude

2. Press F1 (Ref.Level).

Ref.Level

-30.0dBm

F 1

3. Enter the value using the

Numerical and Unit keys, Arrow
keys, and Scroll knob.

BK SP

dB

Enter

GHz /

Sec

mSec

MHz /

kHz /

Sec

μ

*Arrow keys and Scroll knob resolution: vertical
scale

Range dBm

−110 to +20 dBm, 0.1dB resolution

dBmV

−63.01 to +66.99 dBmV, 0.01dB resolution

dBuV

−3.01 to +126.99 dBuV, 0.01dB resolution

See page50 for selecting amplitude units.

Display

Reference Level Reading

Reference
level