GSP-830 User Manual

162

Amplitude

meas:refl:unit?

Returns the reference level unit.
Parameter: 1 (dBm), 2 (dBmV), 3 (dBuV)

meas:refl:unit

Sets the reference level unit. Parameter: 1 (dBm), 2 (dBmV), 3
(dBuV)Example: meas:refl:unit:_1 (dBm)

meas:refl?

Returns the reference level in the current unit.
Example: −30 (−30dBm when the unit is dBm)

meas:refl

Sets the reference level in the current unit.
Example: meas:refl:_−30 (−30dBm when in dBm)

meas:refl:scale?

Returns the amplitude scale.
Parameter: 1(10dB/Div), 2(5dB/Div), 3(2dB/Div), 4(1dB/Div)

meas:refl:scale

Sets the amplitude scale. Parameter: 1(10dB/Div), 2(5dB/Div),
3(2dB/Div), 4(1dB/Div) Example: meas:refl:scale_1 (10dB/Div)

meas:refl:exg?

Returns the external gain/loss in dB. Example: −6 (−6dB)

meas:refl:exg

Sets the external gain/loss. Example: meas:refl:exg_−6 (−6dB)

meas:refl:corr:edit Sets the amplitude correction data (frequency, amplitude). Need

to specify the table index and the number of correction.
Example: meas:refl:corr:edit_1_CR_2,100,−40,150,−30 (table
index1, 2 data, 100MHz/−40dB, 150MHz/−30dB)

meas:refl:corr:edit:
delall

Deletes all point in an amplitude correction set. Need to specify
set index. Example: meas:refl:corr:edit:delall_5 (delete set No.5)

meas:refl:corr:on? Returns the activated amplitude correction set.

Parameter: none, 1 to 5 (correction set)

meas:refl:corr:on? Returns the amplitude correction set is active or inactive. Need

to specify the set index. Parameter: on, off
Example: meas:refl:corr:on_1? (set No.1 is activated?)

meas:refl:corr:on Activates the amplitude correction set. Specify the set index.

Example: meas:refl:corr:on_1 (activate set No.1)

meas:refl:corr:off Deactivates

the

amplitude correction set. Specify set index.

Example: meas:refl:corr:off_1 (deactivate set No.1)

meas:inputz?

Returns the input impedance. Parameter: 0 (50 Ω), 1 (75 Ω)

meas:inputz

Sets the input impedance. Parameter: 0 (50 Ω), 1 (75 Ω)
Example: meas:inputz_0 (50Ω)

meas:inputz:cal? Returns the input impedance calibration value in dB.
meas:inputz:cal

Sets the input impedance calibration value in dB.
Example: meas:inputz:cal_5.9 (5.9dB)

Autoset

meas:autoset:run Runs

autoset.

meas:autoset:amp:auto

Sets the autoset amplitude floor setting to auto mode.

meas:autoset:amp:man

Sets the autoset amplitude floor setting to manual mode.
Need to specify the amplitude in dB.
Example: meas:autoset:amp:man_20 (20dB)