Test Results

Power Blazer

377

Summary (FC BERT)

BER

Note: See BER on page 257 for a description of each alarm/error.

BER Threshold is available when Pass/Fail Verdict is enabled (refer to
page 117).

Bit Error Rate/Count graphically displays a meter representing either
the bit error rate or the bit error count depending on the Pass/Fail
Verdict selection (refer to page 117).

When the verdict is enabled, the values under the threshold are
presented in green while the values above are in red.

When the verdict is disabled, the bit error rate is displayed in blue.

The arrow pointer indicates the current received bit error rate/count.

The Pass/Fail verdict is displayed just on top of the meter when
enabled.

Bit Error and Rate/Amount

Manual allows to enter the amount of manual error to be
generated: 1 (default) through 50.

Rate allows the selection of the injection rate for the selected error.

Max Rate generates the selected error to its theoretical maximum
rate.

Inject generates, if the test is running, the amount of bit error
configured.