beautypg.com

Interface – EXFO FTB/IQS-88000 Series Power Blazer for FTB-2/2Pro/500/IQS-600 User Manual

Page 291

background image

Test Results

Power Blazer

279

Alarms/Errors

Interface

Alarms

LOS

1

(Loss Of Signal)

RX: Absence of an input signal or an all-zeros pattern is received.

LOC Lane

2

(Loss Of Clock Lane) is available for parallel interface

RX: The module is unable to lock on the Physical/CAUI/XLAUI link
interface.

Int. RX Fault (Interlink RX Fault) is available with serial CFP

RX: Fault detected at the interlink point between the FTB/IQS-88000
Series and the CFP, an unexpected signal is coming out from the CFP.

Frequency

2

RX: The received signal frequency meets the standard specifications
(green) or not (red).

For Ethernet 10/100/1000M Electrical, 100M Optical, 1GE Optical,
10GE LAN/WAN, or Fibre Channel 1X/2X/4X/8X/10X/16X interface/rate,
the frequency range is ± 100 ppm.

For parallel interfaces:

1. Available for each Optical Lane for parallel interface.
2. Available for each Physical Lane for parallel interface.

Rate

Optic Type

Frequency

OTU3

4x10G

10.7546 Gbit/s ± 20 ppm

OTU3e1

4x10G

11.1427 Gbit/s ± 20 ppm

OTU3e2

4x10G

11.1458 Gbit/s ± 20 ppm

OTU4

10x10G, 4 x 25G

11.181 Gbit/s ± 20 ppm