Test Results

290

FTB/IQS-88000 Series

Alarms/Errors

OTUx

Note: Available for OTU4, OTU3, OTU3e1, OTU3e2, OTU2, OTU2f, OTU1f, OTU2e,

OTU1e, and OTU1.

Alarms

AIS (Alarm Indication Signal) available with serial interfaces only.

RX: Polynomial number 11 (PN-11) is over all OTU frame bits including
FAS and MFAS for at least 3 consecutive 8192 bit-interval.

TX: Generates polynomial number 11 (PN-11) over all OTU frame bits
including FAS and MFAS continuously.

BDI (Backward Defect Indication)

RX: The BDI bit in the SM overhead field (byte 3, bit 5) is “1” for at least
5 consecutive OTU frames.

TX: Generates “1” for the BDI bit in the SM overhead field (byte 3, bit 5)
continuously.

BIAE (Backward Incoming Alignment Error)

RX: The BEI/BIAE bits in the SM overhead field (byte 3, bits 1 to 4) are
“1011” for at least 3 consecutive frames.

TX: Generates “1011” for the BEI/BIAE bits in the SM overhead field
(byte 3, bits 1 to 4) continuously.

IAE (Incoming Alignment Error)

RX: The IAE bit in the SM overhead field (byte 3, bit 6) is “1” for at least
5 consecutive OTU frames.

TX: Generates “1” for the IAE bit in the SM overhead field (byte 3, bit 6)
continuously.