Far-end – EXFO FTB/IQS-88000 Series Power Blazer for FTB-2/2Pro/500/IQS-600 User Manual
Page 349

Test Results
Power Blazer
337
Performance Monitoring
SEP (Severely Errored Period) (G.828 only): A sequence between
3 to 9 consecutive SES. The sequence is terminated by a second which
is not a SES.
SEPI (Severely Errored Period Intensity) (G.828 only): Gives the count
of SEP events in available time, divided by the total available time in
seconds.
Far-End
EFS (Error Free Second): Gives the count of the seconds within which
no error occurred or when a defect is detected on the near-end.
EC (Error Count) (G.821 only): Gives the number of bit errors.
EB (Errored Block) (G.826, G.828, and G.829): Gives the count of
blocks in which one or more bits are in error.
ES (Errored Second): For G.826, G.828, G.829, M.2100 ISM, and
M.2101: Gives the count of the seconds within which one or more
anomalies (FAS (DSn/PDH), EB, etc.) occurred or at least one defect
occurred.
SES (Severely Errored Second)
For G.826, G.828, G.829 and M.2101: Gives the number of seconds
within which anomalies (FAS (DSn/PDH), EB, etc.) are X% (see table
below) or at least one defect occurred. X = 30% for DSn/PDH signals;
see the following table for SONET/SDH signals SES threshold.
OC-1
STS-1e
STM-0
STM-0e
OC-3
STS-3e
STM-1
STM-1e
OC-12
STM-4
OC-48
STM-16
OC-192
STM-64
OC-768
STM-256
Path
30%
30%
30%
30%
30%
30%
Line/MS
15%
15%
25%
30%
30%
30%
Section/RS
10%
30%
30%
30%
30%
30%