MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-41

Note:

Measurement data is recorded in the LIFEPAK 15 monitor/defibrillator Vital Sign Log. For

more information about the Vital Sign Log and its use, see Chapter 7, "Data Management."

Manual Single-Measurement Procedure

The NIBP measurement typically takes 40 seconds to complete. If the measurement is not

completed within 120 seconds, the cuff automatically deflates.

To obtain a manual single measurement:

1. Press

ON

.

2. Select the appropriately-sized cuff.

3. Properly align the cuff artery markings, if present, and apply snugly to the extremity.

4. Connect the tubing to the cuff and to the NIBP port on the monitor.

5. Change the initial inflation pressure, if necessary.

6. Position the extremity in a relaxed and supported position at approximately the same level as

the patient’s heart. Inform the patient that the cuff will inflate and cause a “big squeeze”

around the arm and that the patient’s fingers may tingle.

7. Press

NIBP

to start the measurement, and check that the patient’s arm is not moving. When

the measurement is complete, systolic, diastolic, and mean arterial pressures are displayed.

To cancel a measurement, press

NIBP

again.

Note:

NIBP pulse rate is displayed only when ECG or SpO

2

is not active.

Timer-Controlled Measurement Procedure

When the timer is set, the monitor performs recurring measurements at a fixed interval. When

using timer-controlled measurement, the interval is counted from the start of the measurement

to the start of the next measurement. Choices are

OFF

(factory default),

2

,

3

,

5

,

10

,

15

,

30

, and

60

minutes.

To take a manual measurement between timer-controlled measurements, press

NIBP

. The next

interval is counted from the beginning of the manual measurement.

Figure 4-12 NIBP Measurements and Timer

Countdown timer—displays time
until next measurement

Systolic pressure

Diastolic pressure

Mean arterial pressure (MAP)