MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-61
Monitoring Continuous Temperature
Intended Use
The LIFEPAK 15 temperature monitor is intended for continuous monitoring of body temperature.
Indications
Temperature monitoring is indicated for use in patients who require continuous monitoring of
body temperature.
Contraindications
None known.
Temperature Monitoring Warnings
WARNINGS
INACCURATE READINGS HAZARDS
POSSIBLE INACCURATE TEMPERATURE READINGS
Using temperature probes or cables that are not approved by Physio-Control may
cause improper temperature monitoring performance and invalidate safety agency
certifications. Use only probes and cables that are specified in these operating
instructions.
POSSIBLE INACCURATE TEMPERATURE READINGS
The Measurement Specialties 4400 Series temperature probes must be used with the
adapter cable that is listed on the Physio-Control website. Using other manufacturers’
connector cables may cause the device to perform improperly.
HEALTH HAZARDS
INFECTION HAZARD
The temperature probe is disposable and intended for single-patient use. Do not clean
and reuse temperature probes. Dispose of contaminated waste according to local
protocols.
POSSIBLE STRANGULATION
Carefully route the temperature probe cable to reduce the possibility of patient
entanglement or strangulation.