beautypg.com

Monitoring continuous temperature, Intended use, Indications – Physio-Control LIFEPAK 15 User Manual

Page 105: Contraindications, Temperature monitoring warnings, Monitoring

background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-61

Monitoring Continuous Temperature

Intended Use

The LIFEPAK 15 temperature monitor is intended for continuous monitoring of body temperature.

Indications

Temperature monitoring is indicated for use in patients who require continuous monitoring of

body temperature.

Contraindications

None known.

Temperature Monitoring Warnings

WARNINGS

INACCURATE READINGS HAZARDS

POSSIBLE INACCURATE TEMPERATURE READINGS

Using temperature probes or cables that are not approved by Physio-Control may

cause improper temperature monitoring performance and invalidate safety agency

certifications. Use only probes and cables that are specified in these operating

instructions.

POSSIBLE INACCURATE TEMPERATURE READINGS

The Measurement Specialties 4400 Series temperature probes must be used with the

adapter cable that is listed on the Physio-Control website. Using other manufacturers’

connector cables may cause the device to perform improperly.

HEALTH HAZARDS

INFECTION HAZARD

The temperature probe is disposable and intended for single-patient use. Do not clean

and reuse temperature probes. Dispose of contaminated waste according to local

protocols.

POSSIBLE STRANGULATION

Carefully route the temperature probe cable to reduce the possibility of patient

entanglement or strangulation.