Screen messages – Physio-Control LIFEPAK 15 User Manual
Page 258

B-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Screen Messages
MAINTENANCE DUE
Reminder message that appears at the interval that is set in
Service mode. Message continues to appear until reset or
turned off.
MANUAL MODE DISABLED
Access to Manual mode from AED mode has been restricted.
MOTION DETECTED!/STOP
MOTION!
Motion was detected during ECG analysis.
NIBP AIR LEAK
NIBP cuff applied too loosely or there is a leak in cuff/monitor
pneumatic system.
NIBP CHECK CUFF
NIBP cuff is not connected to patient or device.
NIBP FAILED
NIBP monitor cannot establish zero-pressure reference.
NIBP FLOW ERROR
NIBP pneumatic system is not maintaining stable cuff pressure.
NIBP INITIALIZING
NIBP requested while NIBP module is still initializing.
NIBP MOTION
Patient extremity moved too much for the NIBP monitor to
accurately complete the measurement.
NIBP OVERPRESSURE
NIBP cuff pressure exceeded 290 mmHg.
NIBP TIME OUT
NIBP monitor did not complete a measurement in 120 seconds.
NIBP WEAK PULSE
The monitor did not detect any pulses.
NO SHOCK ADVISED
The defibrillator did not detect a shockable rhythm.
NO SITES DEFINED
Device is attempting to transmit using Bluetooth connection,
but no associated destinations have been defined.
NOISY DATA! PRESS 12-LEAD TO
ACCEPT
Monitor detects excessive signal interference while acquiring
data. Press 12-LEAD to override the message and acquire 12-
lead ECG with noise.
NON-DEMAND
Pacemaker is in Nondemand (asynchronous) mode.
OBTAINING DEVICE NAMES
Device is obtaining names of available Bluetooth-enabled
devices.
PACER FAULT
Internal error detected during pacing.
PACING IN PROGRESS
The requested action is not available because the device is
currently performing pacing.
PACING STOPPED
Pacing has stopped—for example, due to disconnection of
therapy electrodes.
PASSCODE INCORRECT -
TRY AGAIN
Incorrect passcode entered.
PAUSED
The pacing PAUSE button is pressed and held. Current pulses
are applied at reduced frequency while the MA and PPM
settings are maintained.
Table B-1 Summary of Screen Messages (Continued)
MESSAGE
DESCRIPTION