PADDLE ACCESSORY OPTIONS

6

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

6-5

To remove QUIK-COMBO therapy electrodes from the patient:

1. Slowly peel back the therapy electrode from the edge, supporting the skin as shown in

Figure 6-3.

Figure 6-3 Removing Therapy Electrodes from Skin

2. Clean and dry the patient’s skin.

3. When applying new electrodes, adjust the positions slightly to help prevent skin burns.

4. Close the protective cover on the therapy cable connector when the cable is not in use.

Cleaning

QUIK-COMBO electrodes are not sterile or sterilizable. They are disposable and are for a single

patient application. Do not autoclave, gas sterilize, immerse in fluids, or clean electrodes with

alcohol or solvents.

Include daily inspection of the QUIK-COMBO therapy electrode package as part of your defibrillator

test routine. Daily inspection helps ensure that the therapy electrode has not exceeded the

electrode package Use By date and is ready for use when needed. For more information about daily

inspection and testing, see the Operator’s Checklist in the back of this manual.