beautypg.com

General maintenance and testing, Maintenance and testing schedule, Maintaining the equipment – Physio-Control LIFEPAK 15 User Manual

Page 209

background image

MAINTAINING THE EQUIPMENT

10

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

10-3

General Maintenance and Testing

Periodic maintenance and testing of the LIFEPAK 15 monitor/defibrillator and accessories are

important to help prevent and detect possible electrical and mechanical discrepancies. If testing

reveals a possible discrepancy with the defibrillator or accessories, see "General Troubleshooting

Tips" on page 10-18. If the discrepancy cannot be corrected, immediately remove the LIFEPAK 15

monitor/defibrillator from service and contact a qualified service technician. For testing information

regarding accessories, see the accessory operating instructions.

A

MAINTENANCE DUE

message can be set up to appear at selected intervals (3, 6, or 12 months) to

remind you that the LIFEPAK 15 monitor/defibrillator is due for maintenance. The factory default is

OFF

, but it can be activated by service personnel.

An Operator’s Checklist is included in the back of this manual. You may reproduce the checklist

and use it to inspect and test the LIFEPAK 15 monitor/defibrillator. Daily inspection and test is

recommended.

Maintenance and Testing Schedule

Table 10-1 lists the recommended maintenance and testing schedule. This schedule may be used

in conjunction with the internal quality assurance program of the hospital, clinic, or emergency

medical service where the defibrillator is used.

To ensure proper performance of the monitor/defibrillator, inspect and test the power adapter daily

as described in the Operator’s Checklist.

Cables and paddles are a critical part of therapy delivery and suffer wear and tear. Therapy cable

testing as described in the Operator’s Checklist is recommended on a daily basis. The Test Load

ships with the device and is necessary for testing the QUIK-COMBO cable. Physio-Control

recommends replacement of therapy cables every three years to reduce the possibility of failure

during patient use.

The 12-lead ECG cable is a critical part of diagnosis and suffers wear and tear. Inspect the 12-lead

cable as described in the Operator’s Checklist, and test it as described in "Patient ECG Cable

Check" on page 10-6.

Additional periodic preventive maintenance and testing—such as electrical safety tests,

performance inspection, and required calibration—should be performed regularly by qualified

service technicians. For detailed maintenance recommendations for each feature, see the LIFEPAK

15 Monitor/Defibrillator Service Manual.