Appendix – Physio-Control LIFEPAK 15 User Manual
Page 255

APPENDIX
B
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
B-1
Table B-1 Summary of Screen Messages
MESSAGE
DESCRIPTION
12-LEAD ECG UNAVAILABLE
A 12-lead was requested but the necessary ECG data is not
available.
ABNORMAL ENERGY DELIVERY
A discharge occurred when the paddles were shorted together,
when hard paddles did not have adequate contact with the
patient or were discharged in the air, or patient impedance was
out of range. Message may also appear in certain types of
internal faults.
ACCESS DENIED
Three consecutive incorrect passcode attempts were made to
enter Manual mode.
ACQUIRING 12-LEAD
Monitor is acquiring data for 12-lead ECG report.
ACQUIRING SNAPSHOT
A snapshot report of current vital signs has been requested.
ADVISORY MODE-MONITORING
The device is monitoring the patient ECG for a shockable
rhythm.
ADVISORY: SPCO > 10%
SpCO advisory alert activated. SpCO value is greater than 10%.
ADVISORY: SPMET > 3%
SpMet advisory alert activated. SpMet value is greater than 3%.
ALARM APNEA
No valid breath has been detected for 30 seconds.
ALARMS SILENCED
Alarms are silenced. An alert tone with status message ALARMS
SILENCED
occurs periodically as a reminder.
ANALYZING 12-LEAD
The data for 12-lead ECG report is being analyzed.
ANALYZING NOW-STAND CLEAR
The AED is analyzing the patient ECG rhythm.
ATTEMPTING TO TRANSMIT
The device is processing a transmission request.
AUTO NIBP CANCELLED
The automatic initiation of NIBP measurements has been
cancelled.
BATTERY X LOW
The specified battery has a low energy condition.
BLUETOOTH DEVICE NOT FOUND
Bluetooth device has not been detected.
BLUETOOTH UNAVAILABLE
Unable to locate or connect to target device.
CANNOT CHARGE
CHARGE
is pressed and the synchronize source is missing for
synchronized cardioversion, the therapy cable is not connected,
or QUIK-COMBO electrodes are not attached to the therapy
cable.
CANNOT CHARGE BATTERIES
Both batteries are installed, and the device is unable to charge
either battery.
CANNOT CHARGE BATTERY 1
The device is unable to charge the battery in battery well 1.
CANNOT CHARGE BATTERY 2
The device is unable to charge the battery in battery well 2.