Lifepak, 15 monitor/defibrillator operator’s checklist – Physio-Control LIFEPAK 15 User Manual
Page 287
LIFEPAK
®
15 Monitor/Defibrillator Operator’s Checklist
This is a recommended checklist to use to inspect and test this monitor/defibrillator. Daily inspection
and test is recommended. This form may be reproduced.
Unit Serial No:______________________________ Location:_____________________________________
Instruction
Recommended
Corrective Action
Date
Initials
✔ each box after completing
1. Inspect physical condition for:
Foreign substances
Clean the device.
Damage or cracks
Contact a qualified service technician.
2. Inspect power source for:
Broken, loose, or worn battery pins
Contact a qualified service technician.
Damaged or leaking battery
Recycle or discard battery.
Spare battery available
Obtain fully charged spare battery.
Damage to power adapter and cables
Contact a qualified service technician.
3. Inspect ECG cable and cable port for:
Cracking, damage, broken, or bent
parts or pins
Replace ECG cable.
If port is damaged, contact qualified
service technician.
4. Check ECG electrodes and therapy electrodes for:
Use By date
Replace if date passed.
Spare electrodes available
Obtain spare electrodes.
Damaged, opened package
Discard and replace electrodes.
5. With batteries installed, disconnect from power adapter (if using), press ON and observe for:
Momentary illumination of
self-test messages and LEDs, and
speaker beep
If absent, contact a qualified
service technician.
Two fully charged batteries
Replace low battery or charge installed
battery using power adapter.
Service indicator (
)
If illuminated, contact a qualified service
technician.
6. With batteries installed, reconnect power adapter to device and check for:
(If not using a power adapter, go to Step 7.)
Power adapter LED strip is illuminated If absent, check connections.
If problem persists, contact a
qualified service technician.
Auxiliary power LED on device is
illuminated
Battery charging LED on device is
illuminated or flashing
If absent, check batteries. If problem
persists, contact a qualified service
technician.
7. Perform QUIK-COMBO
®
therapy cable check in Manual mode:*
(If this cable is not used with the defibrillator, go to Step 8.)
a.Disconnect and examine cable for
cracking, damage, broken, or bent
parts or pins.
Replace QUIK-COMBO therapy cable.
b.Connect therapy cable to defibrillator
and the Test Load.
If
CONNECT ELECTRODES
,
PADDLES LEADS OFF
,
CONNECT CABLE
, or
ABNORMAL ENERGY
DELIVERY
message appears,
replace therapy cable and repeat check.
If problem continues, remove the
defibrillator from use and contact a
qualified service technician.
c.Select
LEAD
, then
PADDLES
.
d.Select
200 JOULES
and press
CHARGE
.
e.Press (shock)
button.
f. Confirm
ENERGY DELIVERED
message appears.
If message does not appear, replace
therapy cable and repeat check.
g.Remove Test Load from cable and
verify
PADDLES LEADS OFF
appears.**
If absent, contact a qualified service
technician.