LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
Visual Alarms
Alarms are indicated visually by:
• The violated parameter flashes in inverse video with a message in the message
area of the display.
• These visual indications remain on the display until the alarm is corrected. Visual
indication of alarms continue even when the tones have been silenced.
Table A-5 Setup Options Factory Default Settings
MENU
MENU/ITEM
FACTORY DEFAULT SETTINGS
General
Language
(Country Specific)
Code Summary
Long
Trend Summary
Off
Site Number
000
Device ID
“LP15” + last 4 digits of serial number, for example,
LP151234
Auto Log
On
Line Filter
60 Hz
Timeout Speed
30 seconds
Manual mode
Sync After Shock
Off
Pads Default
200 (joules)
Energy Protocol
Inactive
Internal Default
10 (joules)
Voice Prompts
On
Shock Tone
On
Manual Access
Manual / Direct
Set Passcode
0000
AED mode
Energy Protocol
200–300–360
Auto Analyze
Off
Motion Detection
On
Pulse Check
Never
CPR
CPR Time 1
120 seconds
CPR Time 2
120 seconds
Initial CPR
Off
Initial CPR Time
120 seconds
Preshock CPR
Off
CPR Metronome Metronome
On
Adult - No Airway
30:2
Adult - Airway
10:1
Youth - No Airway
15:2
Youth - Airway
10:1