MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-37
Monitoring Noninvasive Blood Pressure
Intended Use
The LIFEPAK 15 noninvasive blood pressure (NIBP) monitor measures blood pressure (BP) using
the oscillometric measurement technique to determine systolic, diastolic, and mean arterial
pressures, and pulse rate. The measurement can be initiated manually or set to recur automatically
at predetermined intervals.
Blood pressure measurements determined using this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the
American National Standard Electronic or automated sphygmomanometers (AAMI SP-10).
NIBP is a tool to be used in addition to patient assessment. Care should be taken to assess the
patient at all times; do not rely solely on the NIBP monitor.
Indications
Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension
and monitoring BP trends in patient conditions such as, but not limited to, shock, acute
dysrhythmia, or major fluid imbalance.
Contraindications
None known.